Specification
Measurement object | Oxygen concentration in gaseous mixture |
Measurement system | Paramagnetic system |
Measurement range | ۰-۱ to 0-25 vol%O2, 3 ranges can be programmed. |
Analog output signal | ۴ to 20 mA DC (load resistance: Max. 550 Ω) |
Contact output | Fail; 1 point, open or closed when error occurs, user configurable. Low auxiliary gas pressure alarm; 1 point, open or closed when error occurs, user configurable. Maintenance status; 1 point, closed during maintenance. Range Answerback or Hi/Lo Alarm; 2 points, normally de-energized. |
Contact input | Remote range switching, calibration start |
Sample Gas Condition | Gas flow; 300 to 800 ml/min. ±۱۰% of a set value (Standard 600 ml/min) Gas pressure; Approx. 7 kPa in Analyzer inlet Gas temperature; 0 to 50℃ |
Auxiliary Gas | N2 gas, Pressure; 350 to 500 kPa |
Calibration Method | Automatic calibration, Semiautomatic calibration, Manual calibration |
Calibration gas | Zero gas; N2 gas Span gas; Dry air |
Auxiliary gas pressure | N2 gas, 350 to 500 kPa (Approx. 35 ml/min) |
Ambient temperature | -۵ to 50℃ |
Installation | Indoor, panel or wall mounting |
Structure | Flameproof (Ex d IIBT4X) |
Power supply | ۹۰ to 127 V AC, 48 to 63 Hz, 50 or 60 Hz |