Specification
Measurement object | Oxygen concentration in gaseous mixture |
Measurement system | Paramagnetic system |
Measurement range | ۰-۵ to 0-25 vol%O2, 3 range can be programmed |
Analog output signal | ۴ to 20 mA DC (load resistance; Max. 550Ω) |
Contact output | Fail; 1 point, open or closed when error occures, user configurable Maintenance status; 1 point, closed during maintenance Range answerback or Hi/Lo alarm; 2 points, normally de-energized |
Contact input | Remote range switching, calibration start |
Calibration method | Automatic calibration, semiautomatic calibration, manual calibration |
Calibration gas | Zero gas; N2 gas Span gas; Dry air |
Auxiliary gas pressure | N2 gas, 180 kPa(Approx. 35 ml/min) |
Sample gas condition | Gas flow; 200 ml/min ±۱۰% Gas temperature; 0 to 50℃ Humidity; No moisture condensation in the flow path or the sensor |
Ambient temperature | -۵ to 55℃ |
Installation | Indoor, panel or wall mounting |
Power supply | ۱۰۰-۱۱۵ V AC, 50/60 Hz or 200-240 V AC, 50/60 Hz |